Commissioning and Implementation of an EPID Based IMRT QA System 'Dosimetry Check' for 3D Absolute Dose Measurements and Quantitative Comparisons to MapCheck

Posted on:2011-04-20Degree:M.S.B.SType:Thesis
University:The University of ToledoCandidate:Patel, Jalpa AFull Text:PDF
GTID:2444390002450618Subject:Nuclear physics and radiation
Abstract/Summary:PDF Full Text Request
The software package "Dosimetry Check" by MathResolutions, LLC, provides an absolute 3D volumetric dose measurement for IMRT QA using the existing Electronic Portal Imaging Device (EPID) mounted on most linear accelerators. This package provides a feedback loop using the patient's treatment planning CT data as the phantom for dose reconstruction. The aim of this work is to study the difference between point, planar and volumetric doses with MapCheck and Dosimetry Check via the use of the EPID and the diode array respectively. Evaluating tools such as point doses at isocenter, 1-D profiles, gamma volume histograms, and dose volume histograms are used for IMRT dose comparison in three types of cases: head and neck, prostate, and lung. Dosimetry Check can be a valuable tool for IMRT QA as it uses patient specific attenuation corrections and the superiority of the EPID as compared to the MapCheck diode array. This helps reduce the uncertainty in dose for less variability in delivery and a more realistic measured vs computed dose verification system as compared to MapCheck.
Keywords/Search Tags:Dose, IMRT QA, Dosimetry check, EPID