Objective To discuss the clinical value of carotid color Doppler ultrasound(CDUS)ã€computerized tomography angiography(CTA) and digital subtractionangiography(DSA) using in extra cranial internal carotid artery stenosis diseaseexamination.Methods The examinations mentioned-above were performed on302patientswho diognosised possibly carotid artery stenosis disease,observe the sensitivityã€specificity〠rate of accuracy and consistency in carotid color Doppler ultrasoundã€CTA and the twe examination united compared with DSA in carotid artery stenosisdisease.Results Using DSA as a golden standard,compare carotid color Dopplerultrasound to DSA,the sensitivity was81.72%, specificity was59.42%, rate ofaccuracy was67.16%, missed diagnosis rate was18.28%and misdiagnosis ratewas40.58%, positive rate was51.7%,negative rate was85.95%,χ~2=83.043,P<0.01,K=0.362,the ROC Area=0.671.Compare CTA toDSA,the sensitivity was74.26%, specificity was65.48%, rate of accuracy was69.41%, missed diagnosis rate was25.74%and misdiagnosis rate was34.52%,positive rate was63.52%,negative rate was75.86%,χ~2=47.581,P<0.01,K=0.391,the ROC Area=0.710.Compare carotid color Doppler ultrasound unite CTAto DSA,the sensitivity was89.62%, specificity was51.61%, rate of accuracy was69.13%, missed diagnosis rate was10.38%and misdiagnosis rate was48.39%,positive rate was61.29%,negative rate was85.33%,χ~2=44.219, P<0.01,K= 0.399.Compare carotid color Doppler ultrasound to CTA, χ~2=50.231,P<0.01,K=0.467.Conclusion It is unsatisfactory correlativity between carotid color Dopplerultrasound and CTA compared individually with DSA in the examination of carotidartery stenosis disease; It is better correlativity between carotid color Dopplerultrasound unite CTA compared with DSA,Carotid duplex ultrasound unite CTAcan be widely used in carotid artery stenosis disease early screening,but they can’treplace the DSA. |